Handpiece Onsite Warranty
Our workmanship has a 3 - 12 month warranty depending on the model
and work performed.
How to maintain your handpieces:
- Remove handpiece from your work station umbilical tubing.
Then clean at connector junction with isopropyl alcohol only
(do not use harsh chemicals or place in any ultrasonic cleaning units).
- Assure that the bur is still installed, dispense 2 or 3 drops of
lubricant into the air drive port, while rotating the bur between your
fingers with a forward and backward motion at 1/4 turns for 5 to 10 seconds.
- Reconnect the handpiece to the umbilcal tubing while assuring that
the piece is covered with some type of bag. Run the handpiece for 30 - 45
seconds to assure the removal of any excessive lubricant from the turbine.
(This should be performed on every handpiece prior to autoclaving.)
- Once again, remove the handpiece from the umbilical tubing and place
it into the autoclave.
- Remove the handpiece from the autoclave and allow it to cool to room
temperature, approximately 30-40 minutes.
- Re-insert the bur into the handpiece, dispense 2 or 3 drops of lubricant
into the air drive port, while rotating the bur between your fingers with a
forward and backward motion at 1/4 turns for 5 to 10 seconds.
- Reconnect the handpiece to the umbilcal tubing while assuring that the
piece is covered with some type of bag. Run the handpiece for 30 - 45 seconds
to assure the removal of any excessive lubricant from the turbine. (This should
be performed on every handpiece prior to use.)
There are several lubricants in use throughout the many practices and
they are at the discretion of the user. However, these should be used
sparingly to prevent excessive temperatures resulting from residual oils
during highspeed operation in excess of 400,000 RPM.
|
|